Improvement of the stability for determination of low lead content in TSP by FAAS

dc.contributor.authorXiong Xiao-hong
dc.coverage.spatialBolivia
dc.date.accessioned2026-03-22T17:28:56Z
dc.date.available2026-03-22T17:28:56Z
dc.date.issued2010
dc.description.abstractThe instrument parameters are optimized for determination of lead content of less than 50μg/g in TSP by FAAS,including the analysis line,lamp current,slit width,combustion height,acetylene flow,V(C_2H_2)/V(Air).After the optimization,the standard deviation of lead determination is less than 4 mg/kg,both the stability and accuracy of determination are improved.