Calibration and validation of the mini-fiber EC image analysis instrument mean fiber diameter through direct or primary measurements

dc.contributor.authorEduardo Frank
dc.contributor.authorMichel Victor Hubert Hick
dc.contributor.authorMVH Castillo
dc.contributor.authorGaut MC
dc.contributor.authorRubén Herberht Mamani-Cato
dc.coverage.spatialBolivia
dc.date.accessioned2026-03-22T16:25:19Z
dc.date.available2026-03-22T16:25:19Z
dc.date.issued2021
dc.descriptionCitaciones: 1
dc.description.abstractThe Minifiber EC (MFEC) is a portable instrument for measuring the diameter of animal fibers. Its accuracy and precision have been estimated but by comparing its measurements with those of laboratory devices that had been calibrated on other devices in turn, not on a direct or primary measure of diameter. This work attempts to test direct measurements by gravimetry, Vernier mini caliper, microscope and the classic microprojector, using a non-deformable, high resistance synthetic fiber (Kevlar) for direct measurement. The MFEC instrument is calibrated with each mean fiber diameter obtained in direct measurements and its results are compared. The conclusions drawn are that it is possible to calibrate the MFEC instrument with direct measurements on Kevlar and measurement accuracy or tolerance of 0.28 microns is obtained. This indicates a very low biased mean fiber diameter measurement by MFEC.
dc.identifier.doi10.15406/jteft.2021.07.00290
dc.identifier.urihttps://doi.org/10.15406/jteft.2021.07.00290
dc.identifier.urihttps://andeanlibrary.org/handle/123456789/58140
dc.language.isoen
dc.publisherMedCrave Group
dc.relation.ispartofJournal of Textile Engineering & Fashion Technology
dc.sourceNational University of La Rioja
dc.subjectCalipers
dc.subjectCalibration
dc.subjectMaterials science
dc.subjectFiber
dc.subjectOptics
dc.subjectAccuracy and precision
dc.subjectVernier scale
dc.subjectKevlar
dc.subjectMicroscope
dc.subjectMeasuring instrument
dc.titleCalibration and validation of the mini-fiber EC image analysis instrument mean fiber diameter through direct or primary measurements
dc.typearticle