Comparison in calibration of mass and conventional mass SIM.M.M-S11

dc.contributor.authorEL O Becerra
dc.contributor.authorLuis Manuel Peña
dc.contributor.authorBoris Escalante
dc.contributor.authorL Cori
dc.contributor.authorAldo Quiroga
dc.contributor.authorAlfredo Bermúdez
dc.contributor.authorJose Escobar
dc.contributor.authorW Naula
dc.contributor.authorArnaldo Florencio
dc.contributor.authorM L Valenzuela
dc.coverage.spatialBolivia
dc.date.accessioned2026-03-22T17:55:16Z
dc.date.available2026-03-22T17:55:16Z
dc.date.issued2018
dc.description.abstractThe results of the supplementary comparison in calibration of mass and conventional mass SIM.M.M-S11 are reported. This comparison was organized by seven mass laboratories members of Inter-American Metrology System SIM and stainless steel standard weights with nominal values 2kg, 1kg, 200g, 50g, 1g and 200mg were chosen. These values link this comparison with the key comparisons CCM.M-K4 and CCM.M-K5. For the evaluation of the reference value and the associated uncertainty the method of least squares has been chosen, and the determination of the degrees of equivalence was calculated using the normalized deviation criteria. The results show a general consistency of the measurements in mass and conventional mass between the participant laboratories.
dc.identifier.doi10.1088/1742-6596/1065/8/082012
dc.identifier.urihttps://doi.org/10.1088/1742-6596/1065/8/082012
dc.identifier.urihttps://andeanlibrary.org/handle/123456789/67039
dc.language.isoen
dc.publisherIOP Publishing
dc.relation.ispartofJournal of Physics Conference Series
dc.sourceInstituto Nacional de Metrologia, Qualidade e Tecnologia
dc.subjectCalibration
dc.subjectEquivalence (formal languages)
dc.subjectConsistency (knowledge bases)
dc.subjectMetrology
dc.subjectStandard deviation
dc.subjectStandard uncertainty
dc.subjectMathematics
dc.subjectAnalytical Chemistry (journal)
dc.titleComparison in calibration of mass and conventional mass SIM.M.M-S11