Improvement of the stability for determination of low lead content in TSP by FAAS

Date

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

The instrument parameters are optimized for determination of lead content of less than 50μg/g in TSP by FAAS,including the analysis line,lamp current,slit width,combustion height,acetylene flow,V(C_2H_2)/V(Air).After the optimization,the standard deviation of lead determination is less than 4 mg/kg,both the stability and accuracy of determination are improved.

Description

Citation

DOI